中文摘要: |
采用盐酸为介质,以顺序注射-氢化物发生原子荧光法测定生产用水中的锑含量.通过正交实验确定了仪器工作参数,研究了介质浓度、KBH4浓度、硫脲-抗坏血酸加入量等因素对测定结果的影响,探讨了共存离子的干扰.在选定的最佳实验条件下,方法的检出限为0.011ng/mL,线性范围0~8 ng/mL,相对标准偏差为2.78%~6.20%,加标回收率为91.0%~99.1%.该方法操作简便,灵敏度高,准确度好,与微波消解法相比较,两者无显著性差异. |
英文摘要: |
Using HCl as the medium, the determination of antimony in production water by sequence injection- hydride generation atomic fluorescence spectrometry has been reported. Operating conditions of the instrument were evaluated by orthogonal experiments. Effects of acidity, KBH4 concentration, and thiourea-ascorbic acid content, etc. on measurement results were studied. The effect of coexistent elements was also investigated. Under optimum conditions, the detection limit was 0.011 ng/mL with a linear range of 0~8 ng/mL. The relative standard deviation was 2.78%~6.20%, the recovery rate was 91.0%~99.1%. The method is simple, sensitive and accurate. No significant difference appears between this method and the microwave digestion method. |