异硫氰酸荧光素荧光猝灭法测定痕量碘酸根
A Fluorescence Quenching Method for the Determination of Iodate with Fluorescein Isothiocyanate
投稿时间:2000-08-08  修订日期:2000-09-14
DOI:
中文关键词:  碘酸根  异硫氰酸荧光素    荧光猝灭法
英文关键词:iodate  fluorescein isothiocyanate  iodine  fluorescence quenching method
基金项目:
作者单位
朱展才 漳州师范学院化学系, 福建漳州363000 
黄佳英 漳州师范学院化学系, 福建漳州363000 
王佩珊 漳州师范学院化学系, 福建漳州363000 
摘要点击次数: 1121
全文下载次数: 2
中文摘要:
      碘酸根与碘离子反应生成游离碘使异硫氰酸荧光素的荧光猝灭,λex为487nm,λem为517nm,碘酸根含量在2~8μg/L范围内与荧光猝灭程度有良好线性关系,检测限为0.74μg/L.该法灵敏度高,操作简便,用于含碘食盐中微量碘酸根的测定,结果满意.
英文摘要:
      A highly sensitive and simple fluorescence quenching method has been developed to determine the iodate(IO3-)with fluorescein isothiocyanate as the fluorescent reagent (λex=487 nm,λem=517 nm).IO3- reacts with potassium iodide giving iodine(I2)which then reacts with fluorescein isothiocyanate to produce a non-fluo-rescent compound,thus the fluorescence intensity of the system is quenched.The linear calibration range is 2~8μg/L.The detection limit for IO3- is 0.74μg/L.The proposed method has been used to determine iodate in table salts with satisfactory results.
HTML  查看全文  查看/发表评论  下载PDF阅读器